German French Italian Portuguese Japanese Korrean Russian
Home

News

Vertical flow table user guide: Routine maintenance and operation points

Vertical flow table user guide: Routine maintenance and operation points

In many fields, such as scientific research and medical treatment, the vertical flow table is our right-hand man, providing us with a clean working environment. But just like we take good care of the car, the vertical flow table also needs proper daily maintenance and correct operation to maintain a good "working state," and we will discuss this in detail below.
1. Key points of daily maintenance
(1) Cleaning work
Countertop cleaning: Wipe the countertop with a clean, damp cloth before and after each use. If you encounter stubborn stains, do not use strong corrosive cleaning agents at will. Choose a mild, no-residue special cleaning agent, gently wipe, and then wipe with water and dry.
Filter cleaning: This is a top priority! The filter should be checked regularly, generally every 1-2 weeks. If it is dirty, take it down and rinse it with water, dry it, and then put it back. Although the high-efficiency filter does not need frequent cleaning, it is necessary to regularly check its integrity and filtration effect, and once it is found to be damaged or the filtration efficiency is reduced, it must be replaced in time.
Internal cleaning: the inside of the workbench can not be ignored, every once in a while (such as a month), wipe and disinfect the inside with disinfectant, corners, and other places that are easy to hide dirt more carefully.
(2) Equipment inspection
Fan inspection: Before starting every day, listen to the sound of the fan operation is normal, there is no abnormal noise or vibration. If you find a problem, stop the machine immediately to check, it may be a loose fan blade or motor failure, do not drag, so as not to affect the work.
Electrical system inspection: Regularly check whether the power cord has signs of damage and aging and whether the plug and socket are in good contact. It is also necessary to check whether the buttons and indicators on the control panel work normally to ensure the electrical safety of the equipment.
Second, operation precautions
(1) Start preparation
Environmental inspection: Before starting, look at the environment around the workbench, to ensure that there is no debris accumulation around it and that it well well-ventilated.
Equipment preparation: After turning on the power switch of the equipment, do not rush to use it immediately, let the equipment run no-load for 10-15 minutes, and then start to work when the airflow inside it is stabilized and the purification effect reaches the best state.
(2) The working process
Item placement: Place the items to be used neatly on the work table, taking care not to block the outlet and return the air outlet to keep the airflow smooth.
Operation specification: When working, try to reduce the large movement of the arm in the working area to avoid excessive airflow disturbance. And to avoid the severe vibration operation on the workbench, so as not to affect the purification effect.
(3) Shutdown steps
After work: first, clean the items on the work table, and then wipe the table with disinfectant.
Turn off the device: Turn off the power switch and fan switch of the device one by one. If it is not used for a long time, it is best to unplug the power plug, which is both safe and energy-saving.
The daily maintenance and correct operation of the vertical flow table is not complicated, as long as we develop good habits and strictly follow the requirements, it can always escort our work and provide us with a reliable and clean working environment.