       Installation method of ultra-clean bench:
  1. This equipment should be installed in the purification plant or sealed air-conditioning plant at all levels.
  2. The installation place of this equipment should be far away from the high-speed dust source and vibration source.
  3. After installation, the internal equipment and its surrounding environment must be cleaned carefully with ultra-clean hooves or tools that do not produce fibers.
  4. After installation and cleaning, the average wind speed of the purified working area can be measured by a QDF-2 hot ball-type electric anemometer. When the average wind speed <0.32m/s, adjust the regulator voltage and increase the fan input voltage; Conversely, when the average wind speed >0.48m/s, adjust the regulator voltage and reduce the fan input voltage, until the average wind speed to maintain the ideal range of 0.4m/s ± 20%, can be used.
       The use of an ultra-clean bench:
  1. Before use, the ultraviolet germicidal lamp should be turned on 30 minutes in advance to deal with the microorganisms accumulated on the surface of the table in the purification work area, and then turn off the germicidal lamp and start the blower 20 minutes later.
  2. Do not store unnecessary items on the table surface to keep the clean airflow in the work area without interference.
  3. Regularly (generally once every two months) with a QDF-2 hot ball type electric anemometer to measure the average wind speed in the work area, such as found to be inconsistent with the requirements, you can adjust the fan power supply voltage, so that the worktable is in the best working condition.
  4. The ultra-clean worktable can be operated by one person (single-person single-sided worktable) or two persons (single-person double-sided worktable).